Back to Product Guide - Process Instruments
4 3/4 DIGIT PROCESS METER WITH BCD OUTPUT
MODEL 2004-FS - OPTION AVG/RMS
Made in the USA
Electromagnetic Compatibility
Year 2000 Compliant

Price List
Mechanical Reference
NEMA-SIZE Case
Mechanical Reference
Connections Diagram
STANDARD FEATURES
Parallel BCD output
±39,999-count display span
Voltage ranges from ±0.8 V to ±100 V
Current ranges ±1.0 mA to ±50 mA
1500 Vp CMV, 120 dB CMR
60 dB NMR
Bright, 0.56 in (14.2 mm) LED display
115/230 V ac power
1/8 DIN case

OPTIONS
±49,999-count display span
Isolated, stored, three-state parallel BCD
IEEE-488 (GPIB) interface
Digital setpoint controller
Digital peak or valley hold
5 V dc power
Screw-terminal barrier strip
NEMA-4 splash-proof lens cover
NEMA-size case

BCD OUTPUT STANDARD

Model 2004-FS is a process meter with exceptionally high resolution, accuracy and stability. The standard display span is ±39,999 counts, providing a resolution of 0.0025% of full scale per count. An extended display span of ±49,999 counts is optional. Accuracy is up to 99.99% of reading. Span tempco is better than ±50 ppm of reading/°C, as made possible by precision wirewound resistors.

The 2004-FS offers extensive zero and span adjustment capability plus programmable decimal points for use with all popular process signals, such as 4-20 mA, 10-50 mA, 1-5 V and 0-10 V. It also provides exceptional noise rejection. Its conversion circuit uses a crystal oscillator, which is set for either 50 or 60 Hz rejection. For additional display stability, the meter main assembly can be electrically configured for least-significant digit round-off by 2's or 5's.


STANDARD BCD OUTPUT
Unbuffered, non-isolated parallel BCD output is a standard feature of the 2004 and is implemented on the main circuit board. This output is good for about 1 ft and allows the 2004-FS to be interfaced to a digital comparator (Model 872A) or a printer (Model 820A). Additional data and control outputs can be provided by an optional upper board.

OPTIONS
Upper board options include four BCD output boards, an IEEE-488 (GPIB) communications interface, and a single- setpoint controller. Mechanical options for the standard 1/8 DIN case include a screw-terminal barrier strip for signal and power and a splash-proof lens cover which meets NEMA-4 requirements. A NEMA-size case is optional. For additional information, please refer to the Options and Mechanical sections.

SPECIFICATIONS

ANALOG INPUT
VOLTAGE RECEIVER
Most-sensitive scaling
Least-sensitive scaling
Signal Range
±0.8 V
±100 V
Display Range
±39,999 counts
±39,999 counts
Resolution
20 µV/count
2.5 mV/count
Input resistance
> 1.0 Mohm
> 1.0 Mohm
Bias current
1 nA
100 pA
Maximum input
250 Vp
250 Vp

CURRENT RECEIVER
Most-sensitive scaling
Least-sensitive scaling
Signal Range
±1.0 mA
±50 mA
Display Range
±39,999 counts
±39,999 counts
Resolution
50 nA/count
1.25 µA/count
Input resistance
500 ohm
10 ohm
Voltage drop, max
±1.0 V
±1.0 V
Signal power, max
.125 W
.125 W

VOLTAGE & CURRENT RECEIVER
Coarse-span adjustment: Internal attenuator (voltage) or shunt (current)
Fine-span adjustment:
Front-panel accessible potentiometer
Coarse-zero adjustment: Internal resistor divider
Fine-zero adjustment: Front-panel accessible potentiometer
Input configuration: Single-ended


NOISE REJECTION
NMR, SIG HI to SIG GND:
60 dB, 50/60 Hz
CMR, SIG GND to PWR GND: 120 dB
CMV, SIG GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC spacing

ACCURACY AT 25°C
Maximum error:
±0.005% of reading ±3 counts
Span tempco: ±0.005% of reading/°C
Zero drift, max:
±0.01% of reading/°C
Offset drift, max: (±0.01% offset V ±0.01% FS V)/°C
Step response:
1 s to 99.9% of span
Warmup to rated accuracy: 1 hour

ANALOG-TO-DIGITAL CONVERSION
Technique:
Dual-slope, average value
Polarity: Automatic
Signal integration period:
33 ms at 60 Hz, 40 ms at 50 Hz
Read rate: 3.3/s at 60 Hz, 2.8/s at 50 Hz

DISPLAY
Type:
7-segment, red LED
Height: 0.56 in (14.2 mm)
Symbols: ­8.8.8.8.8
Decimal points: Four positions programmable internally or at connector
Overrange indication: Display flashes

DIGITAL INPUTS
Output signals:
TTL-compatible
Input signals: TTL or 5 V CMOS compatible

POWER
AC voltages:
115 V ac ±10% at 60 Hz; 230 V ac ±10% at 50 Hz
DC voltages: 5 V dc ±5%
Power consumption: 4.5 W nominal

ENVIRONMENTAL
Operating temperature:
0 to 50°C
Storage temperature: ­40 to +75°C
Relative humidity: 95% at 40°C (non-condensing)

MECHANICAL
Dimensions:
Newport DIN1A (1/8 DIN) case standard, Newport NEMA-size case optional (see Mechanical section for drawings)
Weight: 17 oz (480 g)
Case material: 94V-0 UL-rated polycarbonate